ELISA Kits

Lablisa® MF-Human IL-6(Interleukin 6) ELISA Kit

Catalog No.: 1PLOHF225
Sensitivity2
Detection Range3
Assay Duration5
Reactivity4

Description

This ELISA kit uses the Sandwich-ELISA principle. The micro ELISA plate provided in this kit has been pre-coated with an antibody specific to Human IL-6. Samples (or Standards) are added to the micro ELISA plate wells and combined with the specific antibody and biotinylated detection antibody specific for Human IL-6. Then Avidin-Horseradish Peroxidase (HRP) conjugate are added successively to each micro plate well and incubated. Free components are washed away. The substrate solution is added to each well. Only those wells that contain Human IL-6, biotinylated detection antibody and Avidin-HRP conjugate will appear blue in color. The enzyme-substrate reaction is terminated by the addition of stop solution and the color turns yellow. The optical density (OD) is measured spectrophotometrically at a wavelength of 450 nm. The OD value is proportional to the concentration of Human IL-6. You can calculate the concentration of Human IL-6 in the samples by comparing the OD of the samples to the standard curve.

Technical Specifications

Catalog Number1PLOHF225
Sensitivity2
Detection Range3
Assay Duration5
Reactivity4

Product Overview

Product name:Lablisa® MF-Human IL-6(Interleukin 6) ELISA Kit
Reactivity:Human
Alternative Names:B-cell differentiation factor, B-cell stimulatory factor 2, BSF2, BSF-2, CDF, CTL differentiation factor , HSF, hybridoma growth factor, IFNB2, IFN-beta-2, IL6, IL-6, Interferon beta-2, interleukin 6 (interferon, beta 2), interleukin BSF-2, interleukin-6, MGI-2A
Assay Type:Sandwich
Sensitivity:0.47 pg/mL
Standard:50pg/mL
Range:0.78-50pg/mL
Sample Type:Serum, Plasma and Tissue homogenate and Other biological samples;Sample volume:25μL
Assay Length:3.5h
Research Area:

Cardiovascular, Cell biology, Neuroscience, Cancer, Immunology, Metabolism


Standard curve



Concentration (ng/mL)ODCorrected OD
100.002.3662.311
50.001.5641.509
25.000.9610.906
12.500.5290.474
6.2500.2740.219
3.1300.1590.104
1.5700.0980.043
0.000.0550.000


Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level were tested 20 times on one plate, respectively.

Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level were tested on 3 different plates, 20 replicates in each plate, respectively.


Recovery

The recovery of spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.

MatrixRecovery rangeAverage
serum(n=8)82-95%88%
EDTA plasma(n=8)95-107%101%
Cell culture media (n=8)94-107%100%


Linearity

Samples were spiked with high concentrations of target proteins and diluted with Reference Standard & Sample Diluent to produce samples with values within the range of the assay.

Matrix1:21:41:81:16
serum(n=5)93-102%81-96%93-106%82-95%
EDTA plasma(n=5)92-101%87-98%85-92%80-96%
Cell culture media (n=5)86-96%93-104%89-97%85-94%

Product Images

Lablisa® MF-Human IL-6(Interleukin 6) ELISA Kit

Graph

Lablisa® MF-Human IL-6(Interleukin 6) ELISA Kit

Graph Image

Documents

Product Manual
Download

Protocol

This ELISA kit uses the Sandwich-ELISA principle. The micro ELISA plate provided in this kit has been pre-coated with an antibody specific to Human IL-6. Samples (or Standards) are added to the micro ELISA plate wells and combined with the specific antibody and biotinylated detection antibody specific for Human IL-6. Then Avidin-Horseradish Peroxidase (HRP) conjugate are added successively to each micro plate well and incubated. Free components are washed away. The substrate solution is added to each well. Only those wells that contain Human IL-6, biotinylated detection antibody and Avidin-HRP conjugate will appear blue in color. The enzyme-substrate reaction is terminated by the addition of stop solution and the color turns yellow. The optical density (OD) is measured spectrophotometrically at a wavelength of 450 nm. The OD value is proportional to the concentration of Human IL-6. You can calculate the concentration of Human IL-6 in the samples by comparing the OD of the samples to the standard curve.